M133C Electric Power/Energy Calibrator
Operation manual
74
:MVOLtage(?)
<CPD> { OFF | ON | 0 | 1 }
:REFerence
:UNIT(?)
<CPD> { W | VA | VAR }
:PULLup(?)
<CPD> { 0 | 1 }
:CONStant(?)
<DNPD>
:MHARmonics
:UNIT(?)
<CPD> { PRMS | PFUN }
:CONFiguration(?)
<CPD> { 1 | 12 | 123 }
:L280(?)
<CPD> { OFF | ON | 0 | 1 }
OUTP[:STAT](?) <CPD> { ON | OFF }
This command activates or deactivates output terminals.
ON - activates the output
OFF - deactivates the output
If query is sent, calibrator returns ON if the output is active or OFF if it is inactive
Example:
OUTP ON - activates the output
OUTP? - the calibrator returns ON or OFF
OUTP:LOWV(?) <CPD> { FLO | GRO }
This command connects or disconnects the Lo terminals of all voltage outputs to/from GND
terminal.
FLOat ungrounds voltage output
s
GROund grounds voltage output
s
If query is sent, calibrator returns GRO when the output is grounded or FLO when floating.
Example:
OUTP:LOWV GRO
OUTP:LOWV? Calibrator returns FLO or GRO
OUTP:LOWC(?) <CPD> { FLO | GRO }
This command connects or disconnects the Lo terminals of all current outputs to/from GND
terminal.
FLOat ungrounds current output
s
GROund grounds current output
s
If query is sent, calibrator returns GRO when the output is grounded or FLO when floating.
Example:
OUTP:LOWC GRO
OUTP:LOWC? Calibrator returns FLO or GRO
OUTP[:PHAS]:UNIT(?) <CPD> { DEG | COS }