M133C Electric Power/Energy Calibrator
MEATEST
7
Operation manual
2.2
Power-on
Before connecting the calibrator to the mains, check the position of the mains voltage selector
located at the rear panel.
Plug one end of the power cord into the connector located at the rear panel and connect the
other end of the power cord into a wall outlet.
Switch on the mains switch located at the rear panel. Flat display is lit.
The calibrator performs internal hardware checks for 5 seconds. Configuration of the
instrument is displayed in the end of this test (connected phase channels).
- phase unit is connected and ready
N.A.
- phase unit is not connected
- phase is connected, but failed
After the tests conclude, the calibrator resets to its reference state, i.e. the following parameters
are set:
Function
AC power Pac Basic
Voltage range
10 V
Set value
10 V
Current range
1 A
Set value
1 A
Phase
0º
Frequency
50 Hz
Output terminals
OFF
GPIB address of the calibrator is factory-preset to 2. This value is valid until the user changes
it.