M133C Electric Power/Energy Calibrator
MEATEST
59
Operation manual
Connect external voltage 10 V to the calibrator input terminals METER INPUT V-
COM. Exact value can be measured by paralel standard voltmeter.
Adjust main value on the display. It should be equal to the value measured by
standard multimeter. Use
,
,
,
cursor buttons, rotary knob or numeric
keyboard to adjust the main value. Confirm new value by pressing Write softkey.
b)
20 mA DC current range calibration
Select Meter from the calibration menu. Press Select softkey to select the first
calibration point, 0 mA.
Make short on calibrators input terminals METER INPUT mA-COM.
Main value presents the calibration point. Leave value 0.000 and press Write
softkey.
Select calibration point 20 mA.
Connect external current source 20 mA to the calibrator input terminals METER
INPUT mA-COM. Exact value can be measured by serial standard ampmeter.
Adjust main value on the display. It should be equal to the value measured by
standard multimeter. Use
,
,
,
cursor buttons, rotary knob or numeric
keyboard to adjust the main value. Confirm new value by pressing Write softkey.
c)
10 kHz frequency range calibration
Select calibration point 10 kHz.
Connect external AC voltage 10 kHz (voltage between 5 to 10 V) to the calibrator
input terminals METER INPUT V-COM. Exact frequency value can be measured
by paralel standard counter.
Adjust main value on the display. It should be equal to the value measured by
standard counter. Use
,
,
,
cursor buttons, rotary knob or numeric keyboard to
adjust the main value. Confirm new value by pressing Write softkey.