M133C Electric Power/Energy Calibrator
MEATEST
119
Operation manual
This query returns the set value of power using standard exponential format. Example: 100.60 W
is returned as 1.002.
Note:
Query switches the calibrator in Energy AC high current mode.
[SOUR]:EACI[:POW]:UNIT(?) <CPD> { W | VA | VAR }
This command selects units for power in the Energy AC high current mode.
If query is sent, calibrator returns actual units in Energy AC high current mode.
Example:
Select unit W for Energy AC high current mode:
EACI:UNIT W
Note:
Command switches the calibrator in Energy AC high current mode.
[SOUR]:EACI:VOLT(?) <DNPD>
This command sets the output voltage value.
<DNPD>
Parameter represents the value of the output voltage expressed in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Voltage 230.5V:
EACI:VOLT 230.5
Note:
Command switches the calibrator in Energy AC high current mode.
[SOUR]:EACI:CURR(?) <DNPD>
This command sets the output current value in Energy AC high current mode.
<DNPD>
Parameter represents the value of the output current expressed in A.
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.001.
Example:
Current 23.05A:
EACI:CURR 23.05
Note:
Command switches the calibrator in Energy AC high current mode.