This command selects units for power in the Power AC mode.
If query is sent, calibrator returns actual units in Power AC mode.
Example:
Select unit W for Power AC mode:
PAC:UNIT W
Note:
Command switches the calibrator in Power AC mode.
[SOUR]:PAC:VOLT(?) <DNPD>
This command sets the output voltage value.
<DNPD>
Parameter represents the value of the output voltage expressed in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Voltage 230.5V:
PAC:VOLT 230.5
Note:
Command switches the calibrator in Power AC mode.
[SOUR]:PAC:CURR(?) <DNPD>
This command sets the output current value in Power AC mode.
<DNPD>
Parameter represents the value of the output current expressed in A.
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.001.
Example:
Current 23.05A:
PAC:CURR 23.05
Note:
Command switches the calibrator in Power AC mode.
[SOUR]:PAC[:CURR]:PHAS(?) <DNPD>
This command sets the phase angle between voltage and current.
<DNPD>
Represents the phase angle between the output voltage and current in degrees
(PHAS:UNIT=DEG), or specified as a value of the power factor (PHAS:UNIT=COS).