MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
102
PIH:VOLT2:PHAS 120
PIH:VOLT3:PHAS 240
Note:
Command switches the calibrator in Power Interharmonic mode.
[SOUR]:PIH:VOLT<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the voltage output in channel <n>. Disabled outputs are OFF
also during OUTP ON.
ON - enables the output
OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled.
Example:
PIH:VOLT1:ENAB ON
Note:
Command switches the calibrator in Power Interharmonic mode.
[SOUR]:PIH:VOLT<n>:IHAR(?) <DNPD>
This command sets the level of interharmonic component for voltage output in channel <n>.
<DNPD>
Represents the voltage in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Interharmonic component 23.05V in channel 1:
PIH:VOLT1:IHAR 23.05
Note:
Command switches the calibrator in Power Interharmonic mode.
[SOUR]:PIH:CURR<n>(?) <DNPD>
This command sets the output current value in channel <n>
<DNPD>
Parameter represents the value of the output current expressed in A.
<n>
Parameter represents the output channel (1, 2 or 3).
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 11.012 V is returned as 1.001.