M133C Electric Power/Energy Calibrator
MEATEST
129
Operation manual
[SOUR]:CAC:FREQ(?) <DNPD>
This command sets the frequency in Current AC mode.
<DNPD>
Parameter represents the value of frequency expressed in Hz.
If query is sent, calibrator returns the value of frequency using standard exponential format.
Example: 50Hz is returned as 5.001.
Example:
Frequency 60Hz:
CAC:FREQ 60
Notes:
Command switches the calibrator in Current AC mode.
[SOUR]:CDC:CURR(?) <DNPD>
This command sets the output voltage value in Current DC mode.
<DNPD>
Parameter represents the value of the output current expressed in A.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 11.012 A is returned as 1.001.
Example:
Current 23.05A:
CDC:CURR 23.05
Note:
Command switches the calibrator in Current DC mode.
[SOUR]:CACI:CURR(?) <DNPD>
This command sets the output current value in Current AC high current mode.
<DNPD>
Parameter represents the value of the output current expressed in A.
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.001.
Example:
Current 23.05A:
CACI:CURR 23.05
Note:
Command switches the calibrator in Current AC high current mode.
[SOUR]:CACI:FREQ(?) <DNPD>