M133C Electric Power/Energy Calibrator
MEATEST
67
Operation manual
Tables of limits
30 V DC Basic range with linearity test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed (mV)
V-DC
30.0 V
12.0
4.8
V-DC
30.0 V
14.0
5.1
V-DC
30.0 V
16.0
5.4
V-DC
30.0 V
18.0
5.7
V-DC
30.0 V
20.0
6.0
V-DC
30.0 V
22.0
6.3
V-DC
30.0 V
24.0
6.6
V-DC
30.0 V
26.0
6.9
V-DC
30.0 V
28.0
7.2
V-DC
30.0 V
30.0
7.5
V-DC
30.0 V
-12.0
4.8
V-DC
30.0 V
-14.0
5.1
V-DC
30.0 V
-16.0
5.4
V-DC
30.0 V
-18.0
5.7
V-DC
30.0 V
-20.0
6.0
V-DC
30.0 V
-22.0
6.3
V-DC
30.0 V
-24.0
6.6
V-DC
30.0 V
-26.0
6.9
V-DC
30.0 V
-28.0
7.2
V-DC
30.0 V
-30.0
7.5
Table I
DC voltage test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed (mV)
V-DC
10.0 V
10.0
2.5
V-DC
10.0 V
-10.0
2.5
V-DC
70.0 V
70.0
17.5
V-DC
70.0 V
-70.0
17.5
V-DC
140.0 V
140.0
35.0
V-DC
140.0 V
-140.0
35.0
V-DC
280.0 V
280.0
70.0
V-DC
280.0 V
-280.0
70.0
Table II
30 V AC Basic range with linearity test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed (mV)
V-AC
30.0 V
12.0
55
4.8
V-AC
30.0 V
14.0
55
5.1
V-AC
30.0 V
16.0
55
5.4
V-AC
30.0 V
18.0
55
5.7
V-AC
30.0 V
20.0
55
6.0
V-AC
30.0 V
22.0
55
6.3
V-AC
30.0 V
24.0
55
6.6
V-AC
30.0 V
26.0
55
6.9
V-AC
30.0 V
28.0
55
7.2
V-AC
30.0 V
30.0
55
7.5
Table III