M133C Electric Power/Energy Calibrator
MEATEST
107
Operation manual
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:CURR<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the current output in channel <n>. Disabled outputs are OFF also
during OUTP ON.
ON - enables the output
OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled.
Example:
PDIP:CURR1:ENAB ON
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:CURR<n>:CHAN(?) <DNPD>
This command sets the level of change (Dip/Swell - t3 period) for current output in channel <n>.
<DNPD>
Represents the current in A.
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.001.
Example:
Dip/Swell change 2.305A in channel 1:
PDIP:CURR1:CHAN 2.305
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:FREQ(?) <DNPD>
This command sets the frequency in Power Dip/Swell mode.
<DNPD>
Parameter represents the value of frequency expressed in Hz.
If query is sent, calibrator returns the value of frequency using standard exponential format.
Example: 50Hz is returned as 5.001.
Example:
Frequency 60Hz:
PDIP:FREQ 60
Notes:
Command switches the calibrator in Power Dip/Swell mode.