Operation manual
[SOUR]:PDIP:TRIG:REP(?) <CPD> { ONE | REP }
This sets automatic repetition of generated Dip/Swell shape.
ONE – trigger starts generation of one Dip/Swell shape
REP – trigger starts repeated generation of Dip/Swell shapes
If query is sent, calibrator returns ONE if repetition is disabled or REP if it is enabled.
Example:
PDIP:TRIG:REP ONE
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:TRIG:STAR
This command starts generation of Dip/Swell shape.
Example:
PDIP:TRIG:STAR
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:EAC:ENER?
This query returns the set value of energy using standard exponential format. Example: 100.60 Ws
is returned as 1.002.
Note:
Query switches the calibrator in Energy AC mode.
[SOUR]:EAC:DEV?
This query returns deviation between tested energy meter end the calibrator. Calculation of
deviation is based on pulses generated by energy meter. Example: 0.612 % is returned as
6.120000e-001.
Note:
Query switches the calibrator in Energy AC mode.
[SOUR]:EAC:POW?
This query returns the set value of power using standard exponential format. Example: 100.60 W
is returned as 1.002.
Note:
Query switches the calibrator in Energy AC mode.