SmartDGA Hardware Manual
Valve Installation
•
32
16.
Place the mounting ring (PN 115-0044-01) over the stainless steel pipe and the Instrument
Side Adapter as shown below.
Note:
The mounting ring is shipped with the SmartDGA instrument packaging.
17.
Remove the plate cover over the IN port on the instrument side adapter by removing the
two screws with a Phillips screwdriver. Save the plate cover and the two screws for use
during the commissioning process.
Note:
Instead of removing both screws, you may instead remove one screw and leave the
plate cover hanging by the other screw to reduce the likelihood of lost hardware.
18.
Check to ensure that the plug is closed. If not, tighten the isolation plug on the Instrument
Side Adapter with an Allen wrench until it is fully closed.
Closed Plug
Summary of Contents for SmartDGA
Page 1: ...HARDWARE MANUAL SmartDGA...
Page 99: ...SmartDGA Hardware Manual Diagrams Drawings 99 9 5 SmartDGA NEMA 4 Enclosure...
Page 100: ...SmartDGA Hardware Manual Diagrams Drawings 100 9 6 Inline Mounting Plate...
Page 102: ...SmartDGA Hardware Manual Diagrams Drawings 102 9 8 Assy Wall Mount SmartDGA EZHubTM...
Page 103: ...SmartDGA Hardware Manual Diagrams Drawings 103...
Page 104: ...SmartDGA Hardware Manual Diagrams Drawings 104 9 9 Assy Panel Mount SmartDGA EZHubTM...
Page 105: ...SmartDGA Hardware Manual Diagrams Drawings 105...
Page 106: ...SmartDGA Hardware Manual Diagrams Drawings 106 9 10 Assy Wall Mount LumaSMART iCoreTM...
Page 107: ...SmartDGA Hardware Manual Diagrams Drawings 107...
Page 108: ...SmartDGA Hardware Manual Diagrams Drawings 108 9 11 Assy Panel Mount LumaSMART iCoreTM...
Page 109: ...SmartDGA Hardware Manual Diagrams Drawings 109...
Page 110: ...SmartDGA Hardware Manual Diagrams Drawings 110 9 12 SmartDGA Instrument Sun Shade...
Page 111: ...SmartDGA Hardware Manual Diagrams Drawings 111...