Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-235
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
:TRIG:TIM1:CLEar
Clears trigger timer 1.
Also see
(on page 6-235)
:TRIGger:TIMer<n>:STARt:OVERrun?
(on page 6-239)
:TRIGger:TIMer<n>:COUNt
This command sets the number of events to generate each time the timer generates a trigger event or is enabled
as a timer or alarm.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1
Usage
:TRIGger:TIMer<n>:COUNt <count>
:TRIGger:TIMer<n>:COUNt?
<n>
Trigger timer number (1 to 4)
<count>
The number of times to repeat the trigger (0 to 1,048,575)
Details
If
count
is set to a number greater than 1, the timer automatically starts the next trigger timer delay
at the expiration of the previous delay.
Set
count
to zero (0) to cause the timer to generate trigger events indefinitely.
If you use the trigger timer with a trigger model, make sure the count value is the same or more than
any count values expected in the trigger model.
Example 1
TRIG:TIM2:COUN 4
Set the number of events to generate for trigger
timer 2 to four.