Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-108
DMM7510-901-01 Rev. B / May 2015
[:SENSe[1]]:<function>:THERmistor
This command describes the type of thermistor.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
5000
Usage
[:SENSe[1]]:<function>:THERmistor <n>
[:SENSe[1]]:<function>:THERmistor DEFault
[:SENSe[1]]:<function>:THERmistor MINimum
[:SENSe[1]]:<function>:THERmistor MAXimum
[:SENSe[1]]:<function>:THERmistor?
[:SENSe[1]]:<function>:THERmistor? DEFault
[:SENSe[1]]:<function>:THERmistor? MINimum
[:SENSe[1]]:<function>:THERmistor? MAXimum
<function>
The function to which the setting applies; see
Functions
<n>
The thermistor type in ohms:
•
2252:
2252
•
5000:
5000
•
10000:
10000
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command is only applicable when the transducer type is set to thermistor.
For the <n> parameter, only 2252, 5000, or 10000 are valid entries. Other values you enter will cause
an out of range error message. The only exception to this is if you enter 2200 or 2250, the Model
DMM7510 will accept the entry but change it to 2252.
Example
FUNC "TEMP"
TEMP:TRAN THER
TEMP:THER 2252
Set measurement function to temperature.
Set the transducer type to thermistor.
Set the thermistor type to 2252.
Also see
[:SENSe[1]]:<function>:TRANsducer
(on page 6-112)
(on page 2-120)