Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 9: Frequently asked questions (FAQs)
DMM7510-901-01 Rev. B / May 2015
9-7
How do I change the command set?
You can change the command set that you use with the Model DMM7510. The remote command sets
that are available include:
•
SCPI: An instrument-specific language built on the SCPI standard.
•
TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
You cannot combine the command sets.
As delivered from Keithley Instruments, the Model DMM7510 is set to work with the Model
DMM7510 SCPI command set.
To set the command set from the front panel:
1. Press the
MENU
key.
2. Under System, select
Settings
.
3. Select the button next to Command Set.
4. Select the command set.
5. You are prompted to reboot.
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
Why can't the Model DMM7510 read my USB flash drive?
Verify that the flash drive is formatted with the FAT file system. The Model DMM7510 only supports
FAT drives.
In Microsoft
®
Windows
®
, you can check the file system by checking the properties of the USB flash
drive.