Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-183
dmm.measure.rel.method
This attribute determines if relative offset is applied to the measurements before calculating the DC voltage ratio
value.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
dmm.METHOD_PARTS
Usage
value
= dmm.measure.rel.method
dmm.measure.rel.method =
value
value
The method used:
•
Do not apply relative offset:
dmm.METHOD_RESULT
•
Apply relative offset:
dmm.METHOD_PARTS
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command determines if relative offset is applied to the voltage measurements before the ratio
calculation or if the relative offset is applied to the final calculated value.
When the parts methods is selected, the individual readings each have the relative offset value
applied before being used to calculate the measurement reading. When parts is selected, the relative
offset value is working with smaller ranges, so an error may occur. Reduce the relative offset value if
you receive an error. A relative offset is applied to the sense value and then to the input value.
When the results method is selected, the individual readings do not have the relative offset value
applied. The relative offset value is applied to the final calculation.
Example
dmm.measure.func = dmm.FUNC_DCV_RATIO
dmm.measure.rel.method = dmm.METHOD_PARTS
Set the measure function to DC voltage
ratio.
Set the method to apply relative offset
before generating the ratio.
Also see
(on page 8-180)
(on page 8-181)