Example 2
TRACe:MAKE "bufferVolts", 100
TRACe:POINts? "bufferVolts"
TRACe:DELete "bufferVolts"
TRACe:MAKE "bufferVolts", 1000
TRACe:POINts?
Create a buffer named
bufferVolts
to store 100
readings.
Query the size of
bufferVolts
.
Output:
100
Delete the buffer named
bufferVolts
.
Make a new buffer named
bufferVolts
to store 1000
readings.
Query the size of
bufferVolts
again to verify it can
store 1000 readings.
Output:
1000
Example 3
TRACe:POINts 5000, "bufferVolts"
TRACe:POINts?
Resize an existing buffer named
bufferVolts
to store
5000 readings.
Query the size of
bufferVolts
to verify it can store
5000 readings.
Output:
5000