Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-57
[:SENSe[1]]:<function>:ATRigger:EDGE:SLOPe
This command defines the slope of the analog trigger edge.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
RISing
Usage
[:SENSe[1]]:<function>:ATRigger:EDGE:SLOPe <setting>
[:SENSe[1]]:<function>:ATRigger:EDGE:SLOPe?
<function>
The function to which the setting applies; see
Functions
<setting>
The direction:
•
Rising:
RISing
•
Falling:
FALLing
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This is only available when the analog trigger mode is set to edge.
Rising causes an analog trigger event when the analog signal trends from below the analog signal
level to above the level.
Falling causes an analog trigger event when the signal trends from above to below the level.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE EDGE
CURR:ATR:EDGE:LEV 2.5
CURR:ATR:EDGE:SLOP RIS
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to edge.
Set the analog trigger level to 2.5 A.
Set the analog trigger slope to rising.
Also see
[:SENSe[1]]:<function>:ATRigger:EDGE:LEVel
(on page 6-56)
[:SENSe[1]]:<function>:ATRigger:MODE
(on page 3-64)