Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-122
DMM7510-901-01 Rev. B / May 2015
Example
:TRAC:CLEAR
:COUN 10
:MEAS?
:TRAC:DATA? 1,10
Clear data from the reading buffer.
Set the count to 10.
Make ten measurements.
Returns the last measurement.
Example output:
-5.693831E-05
Read all ten measurements.
Example output:
-7.681046E-05,-2.200288E-04,-
9.086048E-05,-6.388056E-05,-
7.212282E-05,-4.874761E-05,-
4.741654E-04,-6.811028E-05,-
5.110232E-05,-5.693831E-05
Also see
(on page 6-4)
(on page 6-155)
(on page 6-231)
[:SENSe[1]]:DIGitize:COUNt
This command sets the number of measurements to digitize when a measurement is requested.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
10,000
Usage
[:SENSe[1]]:DIGitize:COUNt <n>
[:SENSe[1]]:DIGitize:COUNt DEFault
[:SENSe[1]]:DIGitize:COUNt MINimum
[:SENSe[1]]:DIGitize:COUNt MAXimum
[:SENSe[1]]:DIGitize:COUNt?
[:SENSe[1]]:DIGitize:COUNt? DEFault
[:SENSe[1]]:DIGitize:COUNt? MINimum
[:SENSe[1]]:DIGitize:COUNt? MAXimum
<n>
The number of measurements (1 to 55,000,000)
Details
The digitizer makes the number of readings set by this command in the time set by the sample rate.
This command does not affect the trigger model.
Example
DIG:COUN 10
Make ten measurements.
Also see
(on page 6-7)