Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-298
DMM7510-901-01 Rev. B / May 2015
Trigger events
Event description
Event constant
Trigger event blender
N
(1 to 2), which combines trigger
events
trigger.EVENT_BLENDER
N
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Analog trigger
trigger.EVENT_ANALOGTRIGGER
External in trigger
trigger.EVENT_EXTERNAL
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist()
command to view the trigger model blocks in a list format.
The following usage has been deprecated; replace it with the usage above that includes the
clear
parameter.
trigger.model.load("LoopUntilEvent",
triggerEvent
,
position
)
trigger.model.load("LoopUntilEvent",
triggerEvent
,
position
,
delay
)
trigger.model.load("LoopUntilEvent",
triggerEvent
,
position
,
delay
,
bufferName
)
trigger.model.load("LoopUntilEvent",
triggerEvent
,
position
,
delay
,
bufferName
,
readingBlock
)
Example
reset()
-- Set up measure function
dmm.measure.func = dmm.FUNC_DC_CURRENT
-- Initiate readings
trigger.model.load("LoopUntilEvent", trigger.EVENT_DISPLAY, 50)
trigger.model.initiate()
Reset the instrument.
Set the instrument to measure current.
Load the LoopUntilEvent trigger model to make measurements until the front panel trigger key is pressed,
then continue to make measurements equal to 50 % of the reading buffer size.
Start the trigger model.
Also see
None