Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 3: Functions and features
DMM7510-901-01 Rev. B / May 2015
3-19
Using the front panel to set buffer capacity:
When you resize a reading buffer, data in the buffer is cleared.
1. Press the
MENU
key.
2. Under Measure, select
Reading Buffers
. The MANAGE READING BUFFERS window is
displayed.
Figure 110: MANAGE READING BUFFERS window
3. Select a reading buffer from the list. For example, select
testData
. The settings for
testData
are displayed.
Select the
Capacity
button and enter the new size for the buffer.
4. Select
OK
. The MANAGE READING BUFFERS window is displayed.
5. Press the
HOME
key to return to the Home screen.
Using SCPI commands to set buffer capacity:
To set the
testData
reading buffer to hold 300 readings, send the following command:
TRACe:POINts 300, "testData"
Using TSP commands to set buffer capacity:
To set the
testData
reading buffer to hold 300 readings, send the following command:
testData.capacity = 300