Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-159
:TRACe:LOG:STATe
This command indicates if information events are logged when the specified reading buffer is at 0 % or 100 %
filled.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
defbuffer1: 1 (ON)
defbuffer2: 1 (ON)
User-created buffer: 0 (OFF)
Usage
:TRACe:LOG:STATe <logState>
:TRACe:LOG:STATe <logState>, "<bufferName>"
:TRACe:LOG:STATe?
:TRACe:LOG:STATe? "<bufferName>"
<
logState
>
Do not log information events:
OFF
or
0
Log information events:
ON
or
1
<
bufferNam
e>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
Details
If this is set to on, when the reading buffer is cleared (0 % filled) or full (100 % filled), an event is
logged in the event log. If this is set to off, reading buffer status is not reported in the event log.
Example
TRACe:LOG:STATe?
Query the log state of
defbuffer1
.
Output:
1
Indicates that the log state is on.
Also see
(on page 3-13)
(on page 3-30)
(on page 2-154)