Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-170
DMM7510-901-01 Rev. B / May 2015
:TRACe:STATistics:PK2Pk?
This command returns the peak-to-peak value of all readings in the reading buffer.
Type
Affected by
Where saved
Default value
Query only
Recall settings
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRACe:STATistics:PK2Pk?
:TRACe:STATistics:PK2Pk? "<bufferName>"
<bufferName>
A string that indicates the reading buffer; the default buffers
(
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no
buffer is specified,
defbuffer1
is used
Example
TRACe:STAT:PK2Pk?
Returns the peak-to-peak reading value in the default
buffer
defbuffer1
.
TRACe:STAT:PK2Pk? "testData"
Returns the peak-to-peak reading value in the user-
defined buffer
testData
.
Also see
(on page 3-13)
(on page 3-30)
(on page 6-166)
(on page 6-167)
(on page 6-168)
(on page 6-169)
(on page 6-170)
:TRACe:STATistics:STDDev?
This command returns the standard deviation of all readings in the buffer.
Type
Affected by
Where saved
Default value
Query only
Recall settings
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRACe:STATistics:STDDev?
:TRACe:STATistics:STDDev? "<bufferName>"
<bufferName>
A string that indicates the reading buffer; the default buffers
(
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no
buffer is specified,
defbuffer1
is used