Section 3: Functions and features
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
3-6
DMM7510-901-01 Rev. B / May 2015
Setting a relative offset value
You can set a specific relative offset value using the front panel or remote commands.
Using the front panel:
1. Press the
FUNCTION
key and select the measure function.
2. Press the
MENU
key.
3. Select
Calculations
.
4. For Rel, select
On
.
5. Select the button next to Rel Value.
6. Enter the value and select
OK
.
Over a remote interface:
•
SCPI commands: Refer to
[:SENSe[1]]:<function>:RELative
(on page 6-92) and
[:SENSe[1]]:<function>:RELative:STATe
(on page 6-96).
•
TSP commands: Refer to
(on page 8-181) and
(on page 8-180).
Using SCPI commands:
Send the commands:
:SENSe:FUNCtion "VOLTage"
:SENSe:VOLTage:RELative <n>
:SENSe:VOLTage:STATe ON
Where
<n>
is the amount of the offset.
To set the relative offset for another function, replace
VOLTage
with
CURRent
or
RESistance
.
Using TSP commands:
Send the commands:
dmm.measure.func = dmm.FUNC_DC_VOLTAGE
dmm.measure.rel.level =
relValue
dmm.measure.rel.enable = dmm.ON
Where
relValue
is the relative offset value.
To set the relative offset for another function, replace
dmm.FUNC_DC_VOLTAGE
with
dmm.FUNC_DC_CURRENT
or
dmm.FUNC_RESISTANCE
.