Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-171
dmm.measure.math.percent
This attribute specifies the reference constant that is used when math operations are set to percent.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
1
Usage
reference
= dmm.measure.math.percent
dmm.measure.math.percent =
reference
reference
The reference used when the math operation is set to percent; the range is -1e12 to
+1e12
Details
The percent math function displays measurements as percent deviation from a specified reference
constant. The percent calculation is:
Where:
•
Percent
is the result
•
Input
is the measurement (if relative offset is being used, this is the relative offset value)
•
Reference
is the user-specified constant
Example
dmm.measure.func = dmm.FUNC_DC_VOLTAGE
dmm.measure.math.format = dmm.MATH_PERCENT
dmm.measure.math.percent = 50
dmm.measure.math.enable = dmm.ON
Set the measurement function to voltage.
Set the math operations to percent.
Set the reference constant to 50 for voltage
measurements.
Enable math operations.