Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-226
DMM7510-901-01 Rev. B / May 2015
Details
If the trigger layer is not bypassed, the External In/Out terminal is asserted. The arm layers do not
assert the external in/out terminal.
You can use this template to emulate other trigger models if you use only one of the arm layers. Set
the other arm layers to a source of
NONE
, a count of
1
and a delay of
0
.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
Trigger events
Event description
Event constant
No trigger (immediate)
NONE
Front-panel TRIGGER key press (manual trigger)
DISPlay
Notify trigger block
<n>
(1 to 8) generates a trigger event when
the trigger model executes it
NOTify<n>
A command interface trigger (bus trigger):
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
<n>
(1 to 6)
DIGio<n>
External in trigger
EXTernal
Example
Refer to the application notes for the Model DMM7510 on the
http://www.keithley.com
for an example with additional detail about this command.
Also see
None