Section 3: Functions and features
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
3-84
DMM7510-901-01 Rev. B / May 2015
Front panel options
When you select the Notify block from the front panel, the following options are available.
Setting
Description
Notify
The line or timer that is notified:
•
Digital Out Line: Select the digital output event (line 1
through line 6)
•
TSP-Link Out Line: Select the TSP-Link output event (line 1
through 3)
•
Timer: Select the timer event (timer 1 through 4)
•
External Out
•
LAN Out: Select the LAN output line (line 1 to 8)
Config
The available settings depend on the Notify selection:
•
For the digital, TSP-Link, external, and LAN out lines, select
the pulse logic (negative or positive)
•
For the timer, select either delay or start time
Branching blocks
A branch block goes to a trigger block other than the sequential execution block. For example, if you
need to set up the trigger model to wait for an event under some conditions but not others, you can
use a branch block to define when the wait block is enabled. You can use the Branch Once block to
create a bypass and skip the wait block the first time the trigger model runs. This makes it possible to
avoid deadlock when multiple instruments are being synchronized and each one is waiting for
notification from the other one to start the trigger model.
Loop Counter block
When trigger model execution reaches a loop counter block, it goes to a specified block until the
count value is reached. When the counter exceeds the count value, trigger model execution ignores
the branch, continues to the next block in the sequence, and resets the counter.
The counter is reset to 0 when the trigger model starts. It is incremented each time trigger model
execution reaches the counter block.
If you are using remote commands, you can query the counter. The counter is incremented
immediately before the branch compares the actual counter value to the set counter value. Therefore,
the counter is at 0 until the first comparison. When the trigger model reaches the set counter value,
branching stops and the counter value is one greater than the setting.
When you select the Loop Counter block, the following options are available.
Setting
Description
Target Count
The number of times to repeat
Branch to Block
The block number to execute when the counter is less than the
Target Count value