Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-26
DMM7510-901-01 Rev. B / May 2015
bufferVar.endindex
This attribute indicates the last index in a reading buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
bufferVar
.endindex =
endIndex
endIndex
Ending index of the buffer
bufferVar
The name of the reading buffer, which may be a default buffer (
defbuffer1
or
defbuffer2
) or a user-defined buffer
Details
Use this attribute to find the ending index in a reading buffer.
Example
test1 = buffer.make(100)
dmm.measure.count = 6
dmm.measure.read(test1)
print(test1.startindex, test1.endindex, test1.capacity)
dmm.measure.read(test1)
print(test1.startindex, test1.endindex)
Create a buffer named
test1
with a capacity of 100 readings.
Set the measure count to 6.
Make measurements and store them in buffer
test1
.
Get the start index, end index, and capacity of
test1
.
Output:
1, 6, 100
Make six more measurements and store them in buffer
test1
.
Get the start index and end index of
test1
.
Output:
1, 12
Also see
(on page 8-36)
(on page 8-18)
(on page 3-13)
(on page 3-30)