Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-104
DMM7510-901-01 Rev. B / May 2015
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This specifies which math operation is performed on measurements for the selected measurement
function.
You can choose one of the following math operations:
•
y = mx+b
: Manipulate normal display readings by adjusting the m and b factors.
•
Percent
: Displays measurements as the percentage of deviation from a specified reference constant.
•
Reciprocal
: The reciprocal math operation displays measurement values as reciprocals. The displayed
value is
1/X
, where
X
is the measurement value (if relative offset is being used, this is the measured
value with relative offset applied).
Math calculations are applied to the input signal after relative offset and before limit tests.
Example
dmm.digitize.func = dmm.FUNC_DC_VOLTAGE
dmm.digitize.math.format = dmm.MATH_RECIPROCAL
dmm.digitize.math.enable = dmm.ON
Enables the reciprocal math operation on
digitize voltage measurements.
Also see
Calculations that you can apply to measurements
(on page 3-7)
(on page 8-168)