Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-59
[:SENSe[1]]:<function>:ATRigger:MODE
This command configures the type of signal behavior that can generate an analog trigger event.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
OFF
Usage
[:SENSe[1]]:<function>:ATRigger:MODE <setting>
[:SENSe[1]]:<function>:ATRigger:MODE?
<function>
The function to which the setting applies; see
Functions
<setting>
The setting:
•
Edge (signal crosses one level):
EDGE
•
Pulse (two complementary edge events meet a specified time constraint):
PULse
•
Window (signal enters or exits a window defined by two levels):
WINDow
•
No analog triggering:
OFF
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
When edge is selected, the analog trigger occurs when the signal crosses a certain level. You also
specify if the analog trigger occurs on the rising or falling edge of the signal.
When pulse is selected, the analog trigger occurs when a pulse passes through the specified level
and meets the constraint that you set on its width. You also specify the polarity of the signal (above or
below the trigger level).
When window is selected, the analog trigger occurs when the signal enters or exits the window
defined by the low and high signal levels.
Example
FUNC "CURR"
CURR:RANGE 3
CURR:AZER OFF
CURR:ATR:MODE EDGE
CURR:ATR:EDGE:LEV 2.5
CURR:ATR:EDGE:SLOP RIS
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to edge.
Set the analog trigger level to 2.5 A.
Set the analog trigger slope to rising.
Also see
(on page 3-64)