Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-302
DMM7510-901-01 Rev. B / May 2015
allPattern
The bit pattern that is sent when all limits have failed; 1 to 15; default is 15
startInLine
The input line that starts the test;
5
for digital line 5,
6
for digital line 6, or
7
for external
in; default is 5
startDelay
The delay time before each measurement (167 ns to 10 ks); default is 0 for no delay
endDelay
The delay time after the measurement (167 ns to 10 ks); default is 0 for no delay
bufferName
The name of the reading buffer, which may be a default buffer (
defbuffer1
or
defbuffer2
) or a user-defined buffer; defaults to
defbuffer1
Details
This trigger model template allows you to sort components and place them into up to four bins, based
on the comparison to limits.
To set a limit as unused, set the high value for the limit to be less than the low limit.
All limit patterns and the all fail pattern are sent on digital I/O lines 1 to 4, where 1 is the least
significant bit.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist()
command to view the trigger model blocks in a list format.
Example
For a detailed example, see the section in the
Model DMM7510 User's Manual
named "Grading and
binning resistors."
Also see
None