Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-237
:TRIGger:TIMer<n>:DELay
This command sets and reads the timer delay.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
10e-6 (10
µ
s)
Usage
:TRIGger:TIMer<n>:DELay <interval>
:TRIGger:TIMer<n>:DELay?
<n>
Trigger timer number (1 to 4)
<interval>
Delay interval (8e-6 s to 100,000 s)
Details
Each time the timer is triggered, it uses this delay period.
Reading this command returns the delay interval that will be used the next time the timer is triggered.
Example
TRIG:TIM2:DEL 50E-6
Set trigger timer 2 to delay for 50 µs.
Also see
None
:TRIGger:TIMer<n>:STARt:FRACtional
This command configures an alarm or a time in the future when the timer will start.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0
Usage
:TRIGger:TIMer<n>:STARt:FRACtional <time>
:TRIGger:TIMer<n>:STARt:FRACtional?
<n>
Trigger timer number (1 to 4)
<time>
The time in fractional seconds (0 to < 1 s)