Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-28
DMM7510-901-01 Rev. B / May 2015
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command sets the lower limit for the limit
Y
test for the selected measure function. When limit
Y
testing is enabled, this causes a fail indication to occur when the measurement value is less than this
value.
Default is 0.3 for limit 1 when the diode function is selected. The default for limit 2 for the diode
function is –1.
Example
:CALC2:VOLT:LIM1:CLE:AUTO OFF
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
:CALC2:VOLT:LIM1:CLE
Set limit autoclear off.
Enable the beeper for limit 1 when a voltage
measurement exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results
display on the front panel
Return the test results; example output if the test
fails on the low limit:
LOW
Clear the test results.
Also see
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]