Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-310
DMM7510-901-01 Rev. B / May 2015
The following table shows the constants for the events.
Trigger events
Event description
Event constant
No trigger event
trigger.EVENT_NONE
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Notify trigger block
N
(1 to 8) generates a trigger event when
the trigger model executes it
trigger.EVENT_NOTIFY
N
A command interface trigger (bus trigger):
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
trigger.EVENT_COMMAND
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
N
(1 to 6)
trigger.EVENT_DIGIO
N
Line edge detected on TSP-Link synchronization line
N
(1 to 3)
trigger.EVENT_TSPLINK
N
Appropriate LXI trigger packet is received on LAN trigger
object
N
(1 to 8)
trigger.EVENT_LAN
N
Trigger event blender
N
(1 to 2), which combines trigger
events
trigger.EVENT_BLENDER
N
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Analog trigger
trigger.EVENT_ANALOGTRIGGER
External in trigger
trigger.EVENT_EXTERNAL
Example
trigger.model.setblock(6, trigger.BLOCK_BRANCH_ON_EVENT, trigger.EVENT_DISPLAY,
2)
When the trigger model reaches this block, if the front-panel TRIGGER key has been pressed, the trigger
model returns to block 2. If the TRIGGER key has not been pressed, the trigger model continues to block 7
(the next block in the trigger model).
Also see
(on page 3-88)