DIGISPENSE 3020 CONTROLLER MODULE
Page 3-92
7 April, 2011
IVEK Corp.
P03L011B
I/O Test Screen Description -
The I/O Test screen is used for trouble shooting the inputs and outputs of the
Controller Module. Each input and output can be exercised to verify functionality. The current permission level
determines which push-buttons are available as shown.
The Start push-button is not functional, but the Stop push-button is functional while this screen is visible. Normal
operation ceases for rear panel interfaces being controlled only by this screen.
Trigger In, Input 2, Input 3, Input 4, CC Trig In -
The status of the input signals are displayed here. The status will
be either OFF if no external signal is applied or ON if an external signal is applied.
Ready Out, Fault Out, Config Out 3, Config Out 4, Auxiliary Out -
The status of the output signals are displayed and
changed here. A status of OFF indicates the output is not conducting and a status of ON indicates the output is conducting.
RS232 -
The RS232 In displays data being received through the interface, "<no input>" appears and scrolls off as
characters are received. The text "232 test" flashes on the screen at RS232 Out as the text is sent through the interface.
RS485 -
The RS485 connection has not been configured and is reserved for a future enhancement.
To change a value;
1. Press the Arrow push-buttons to highlight the parameter.
2. Press the Enter push-button to bring up the Select New Value screen.
3. Press the Arrow push-buttons to select the desired value.
4. Press the Enter push-button.