DIGISPENSE 3020 CONTROLLER MODULE
Page 3-33
7 April, 2011
IVEK Corp.
P03L011B
3.2.8 Setup (Figure 3.4)
The following steps will guide you through a basic setup for turning your system on for the first time:
1. On the rear of the Controller Module
a. Make sure the 1/0 (On/Off) (1a) switch is set to 0 (Off).
b. Connect the power cord to the Controller Module (1b) and the power source.
c. Connect the Actuator Cable, the larger connector connects to the Controller Module (1c).
2. On the Actuator Module
a. Connect the other end of the Actuator Cable (2a).
b. Make sure the Actuator is secure.
3. On the Pump Module
NOTE
Make sure the inlet tubing is larger than, or the same size as, the outlet tubing. Start with the reservoir even
with or slightly higher than the Pump Module and the Pump Module even with or slightly higher than the
dispense tip. Adjust as necessary to fit your application.
a. Connect the inlet fitting and tubing (3a)
b. Connect the gland inlet fitting and tubing (3b) (
*
optional)
c. Connect the outlet fitting and tubing (3c)
d. Connect the gland outlet fitting and tubing (3d) (
*
optional)
3.2.9 Start-up
a. Switch the 1/0 power switch to the "1" position.
b. The display will illuminate and show the following information. The status will change from Initializing to Idle
and the information for
your
system will be displayed including the current recipe and pump size and there
may be additional push-button selections depending on your system settings.
Refer to Section 3.2.4 in your manual for
field descriptions.
REFERENCE REQUIRED may be flashing
indicating the system needs to be referenced.
c. Press the Reference push-button to reference
the system and the Actuator Module will
reference.
Go to the Dispense section (3.2.9.2) or Meter section
(3.2.9.3) if the system has previously been primed. If not,
follow the instructions in the following section.
NOTE
All Controller Modules are shipped from the factory
with a permission level of Supervisor.