DIGISPENSE 3020 CONTROLLER MODULE
Page 3-105
7 April, 2011
IVEK Corp.
P03L011B
Command Response
Description
t4
t4,<value2> Returns the current Autotrigger Dwell in seconds.
t4,<value2> t4,<value2> Sets the Autotrigger Dwell.
<value2>:
Maximum:
300.00
Minimum:
0.00
(default)
t5
t5,<value2> Returns the Bubble Intake / Agitate Dwell in seconds.
t5,<value2> t5,<value2> Sets the Bubble Intake / Agitate Dwell .
<value2>:
Maximum:
300.00
Minimum:
0.00
Default:
0.05
v => DISPENSE VOLUME
v or v0
v0,<value2> Returns the Production Mode volume.
v0,<value2> v0,<value2> Sets the Production Mode volume.
<value2>:
Dependent on pump selection and mode.
v1
v1,<value2> Returns the Fluidic Mode volume.
v1,<value2> v1,<value2> Sets the Fluidic Mode volume.
<value2>:
Dependent on pump selection and mode
v2
v2,<value2> Returns the Load Threshold.
v2,<value2> v2,<value2> Sets the Load Threshold.
<value2>:
Dependent on pump selection and Mode
v3
v3,<value2> Returns the Fluidic Mode isolation volume.
v3,<value2> v3,<value2> Sets the Fluidic Mode isolation volume.
<value2>:
Dependent on pump selection and Mode
w => DRAWBACK
w or w0
w0,<value2> Returns the Drawback Volume in microliters.
w0,<value2> w0,<value2> Sets the Drawback Volume.
<value2>:
Minimum = 0
Maximum = Chamber volume of current pump - Dispense Volume
w1
w1,<value2> Returns the current Drawback Rate in microliters/second.
w1,<value2> w1,<value2> Sets the Drawback Rate.
<value2>:
Dependent on pump selection and Mode
w2
w2,<value2> Returns the current Drawback Dwell in seconds.
w2,<value2> w2,<value2> Sets the Drawback Dwell.
<value2>:
Maximum:
2.55
Minimum:
0.00
Default:
0.05
y => MOTOR/BASE
y or y0
y0,<value2> Returns the current Acceleration setting.
y0,<value2> y0,<value2> Sets the Acceleration configuration.
<value2>:
0 = Slow