DIGISPENSE 3020 CONTROLLER MODULE
Page 3-70
7 April, 2011
IVEK Corp.
P03L011B
Auto Trig Screen Description -
The Auto Trig screen displays the information for Auto Trigger and allows a user with
Supervisor permission to change the parameters. The current permission level determines which push-buttons are
available as shown. Refer to section 3.3.1.6 for a description of Auto Trig.
The Start push-button is not functional, but the Stop push-button is functional while this screen is visible.
Status -
The status indicates the operational state of the system. Refer to section 3.3.9 for an explanation of each
status type.
Auto Retrigger -
The Auto Retrigger setting indicates the current mode of the trigger. The mode can either be
Disabled, Count or Infinite. (Refer to section 3.3.1.6)
Retrigger Dwell -
The Retrigger Dwell indicates the amount of time in seconds between a dispense and the next auto
retrigger. The range is 0.0 seconds to 300.0 seconds with an increment value of 0.01. (Refer to section 3.3.1.6)
Dispense Count -
The Dispense Count indicates the total number of dispenses to be triggered. The range is 2 to
65535 with an increment value of 1. (Refer to section 3.3.1.6)
Reload Count -
The Reload Count indicates the number of dispenses between each reload. The range is 0 to 65535
with an increment value of 1. (Refer to section 3.3.1.6)
To change a value;
1. Press the Arrow push-buttons to highlight the parameter.
2. Press the Enter push-button to bring up either the Select New Value or Enter New Value screen.
3. In the Select New Value screen use the up and down arrow push-buttons to select the value. In the Enter New
Value screen press the Minimum or Maximum push-buttons to select the desired value or use the left and right
arrow push-buttons to select the digit and the up and down arrow push-buttons to change the digit's value.
4. Press the Enter push-button.