DIGISPENSE 3020 CONTROLLER MODULE
Page 3-100
7 April, 2011
IVEK Corp.
P03L011B
Command Response
Description
f => REFERENCE
f or f0
f0
References the pump.
g => TOTALIZER
g or g0
g0,<value2> Returns the volume totalizer value in microliters.
g0,<value2> g0,<value2> Resets the volume totalizer to 0.
<value2>: 0 = Resets the totalizer value to zero.
g1
g1,<value2> Returns the cycle totalizer value in number of Cycles.
g1,<value2> g1,<value2> Resets the cycle totalizer to 0.
<value2>: 0 = Resets the totalizer value to zero.
g2
g2,<value2> Returns the last prime volume/agitate volume read in microliters.
MAX VALUE: The totalizer will increment to a maximum value of 3,999,999,999 then 'wrap around’.
h => HARDWIRED READY SIGNAL OPERATION
h or h0
h0,<value2> Returns the Auxiliary Output configuration setting.
h0,<value2> h0,<value2> Sets the Auxiliary Output configuration.
<value2>
0 = Disabled
1 = Conducts during fluid mode
2
= Conducts during both (default)
3
= Conducts during prod mode
h1
h1,<value2> Returns the Controller address for RS485 serial interface setting.
h1,<value2> h1,<value2> Sets the Controller address for RS485 serial interface configuration.
<value2>
0 = Minimum (default)
99 = Maximum
h2,<value2> h2,<value2> Sets the contrast percentage.
<value2>
30 = Minimum
50 = (default)
70 = Maximum
h12
h12,<value2>Returns the Input 2 setting.
h12,<value2>h12,<value2>Sets the Input 2 configuration.
<value2>
0
= Clear fault & reference
1
= Fluidic trigger
2
= Enable motion
9 = Load trigger (default)
10 = Port Park
h13
h13,<value2>Returns the Input 3 setting.
h13,<value2>h13,<value2>Sets the Input 3 configuration.
<value2>
0 = Clear fault & reference (default)
1
= Fluidic trigger
2
= Enable motion
9
= Load trigger
10 = Port Park
h14
h14,<value2>Returns the Input 4 setting.
h14,<value2>h14,<value2>Sets the Input 4
configuration. <value2>