![IVEK DIGISPENSE 3020 Manual Download Page 11](http://html1.mh-extra.com/html/ivek/digispense-3020/digispense-3020_manual_2097276011.webp)
DIGISPENSE 3020 CONTROLLER MODULE
Page 3-11
7 April, 2011
IVEK Corp.
P03L011B
NOTE
References to push-buttons for fields 3 through 6 refer to the push-button located below a word on the
display. For example, the push-button located under 'REFERENCE' will be called the Reference push-button.
These are referred to as Function Push-Buttons 1 through 4 (FPB1 - FPB4).
Field 1 -
This field displays screen name and status
information.
The status indicates the operational state of the system. The
status is displayed in the upper right hand corner of the
display. Refer to section 3.3.9 for additional information.
Field 2 -
This field displays all the operating parameters.
Fields 3 through 6 (Function Push-Buttons 1 - 4) -
These fields provide function legends for the four function
push-buttons directly below the screen.
NOTE
1
2
(FPB1)
(FPB2)
(FPB3)
(FPB4)
3
4
5
6
A legend does not appear if the function is not permitted due to permission levels (section 3.2.6) or the value
of the Serial Interface "k1" command (Table 3.3)
.
3.2.5 Help
Help is available when selecting or entering new values by pressing the FPB1 push-button. Information pertaining to
the selected value will be displayed. Press the FPB1 key again to clear the help information from the screen.
3.2.6 Permission Levels
There are three levels of permission; Operator, I/O Test and Supervisor. Each level allows access to selected items
on the display. The permission level can be changed in the Change Permission screen (Section 3.2.7.8).
The highest level is
Supervisor
. This level allows access to all items on all screens. The
I/O Test
level allows access
to the same items as the Operator except it also allows access to the I/O Test screen. The
Operator
level does not
allow changing system parameters and is the typical use setting. Recipes can be selected and loaded, but not
changed. See serial interface 'k1' command for additional restrictions.
3.2.7 Screens
Sections 3.2.7.1 through 3.2.7.11 provide an image of each screen, the permission level access and a brief
description of each parameter. All screens are described in Section 3.3 - Advanced Operation. For ease of use, the
screen will always be on the left and the description will always be on the right when the two pages are viewed
together.
The following alphabetical list provides a brief description of each screen and figure 3.3 shows the screens and
provides a map of how to navigate to each screen. Transition between screens is limited by permission level,
operating mode and interface signals.