DIGISPENSE 3020 CONTROLLER MODULE
Page 3-35
7 April, 2011
IVEK Corp.
P03L011B
3.2.9.1 Prime
Press the PRIME SCREEN push-button to enter Prime Mode.
Position the pump for priming (outlet above or even with the inlet) fill all reservoirs with liquid, and make sure all liquid
lines are connected and secure. Press the START push-button to begin priming. The system cannot be initiated using
the rear panel "CC TRIG" signal. The system will prime based on the Direction, Volume and Rate settings. Pressing
the STOP push-button can stop the priming at any time.
3.2.9.2 Dispense
Press the DISPENSE SCREEN push-button to enter Dispense Mode. If the DISPENSE SCREEN is not shown go to
SETUP A screen and change Production Mode to DISPENSE.
To start dispensing press the START button or supply a trigger to the rear panel. The system will dispense based on
the Direction, Volume, Rate, Load Rate, Load Threshold and Drawback settings. Pressing the STOP push-button can
stop the dispensing at any time.
When drawback is enabled, the Volume setting specifies the net fluid displaced, the actual forward motion is the sum
of the specified dispense volume and the drawback volume. The Drawback Rate during drawback and the Drawback
Dwell (time between the forward and reverse portions of the cycle) settings are shown on the Drawback screen.
3.2.9.3 Meter
Press the METER SCREEN push-button to enter Meter Mode. If the METER SCREEN is not shown go to SETUP A
screen and change Production Mode to METER.
To start metering press the START button or supply a trigger to the rear panel. The system will meter based on the
Direction, Load Threshold, Rate, Load Rate and Drawback settings. Pressing the STOP push-button can stop the
metering at any time.
When drawback is enabled, the Drawback Rate during drawback and the Drawback Dwell (time between the forward
and reverse portions of the cycle) settings are shown on the Drawback screen.
3.2.10 Faults
Faults cause the Fault screen (section 3.2.7.11) to appear. If a fault occurs, the type of fault will be displayed on the
screen. Once the problem is corrected, press the Reference push-button. Refer to section 3.3.13 for a list and
description of each fault.
3.3
ADVANCEDOPERATION
The Controller Module provides the controls for producing fluid flow via a positive displacement pumping mechanism.
The systems utilize solid-state electronics, stepping motor drives, and precision machined ceramic pump heads.
These components combine to provide exceptional accuracy and precision, high reliability, and low maintenance.
Volume commands for the Controller Module use microliters. Rate commands are in microliters per second. Pumping
is started using the Start push-button
based on the screen being viewed. Push the Stop push-button
to
stop the operation before the respective volume setting is reached.
3.3.1 Production Mode
Production Mode provides precision dispensing and metering operations. Three Dispense Modes (Dispense,
Dispense Minimum Chamber Volume, Dispense Multichamber) and two Meter Modes (Meter, Meter Multichamber)