DIGISPENSE 3020 CONTROLLER MODULE
Page 3-102
7 April, 2011
IVEK Corp.
P03L011B
Command Response
Description
k1,<value2> k1,<value2> Sets the Front Panel Lock configuration.
<value2>
0
= Enable All (default)
1
= Disable Recipe Save
2
= Disable Recipe Save & Value Change
3
= Disable Recipe Save, Value Change & Prime Direction
4
= Disable Recipe Save, Value Change & Recipe Get
5
= Disable Recipe Save, Value Change, Recipe Get & Prime Direction
6
= Disable Recipe Save, Value Change, Recipe Get Prime Direction & Start Stop
k2
k2,<value2> Returns the current Change Permission level.
0 = Operator
1 = I/O Test
2 = Supervisor (default)
k2,<value2> k2,<value2> Change Permission.
<value2>
Password for Desired Permission Level in command
Current Permission Level in response.
Maximum: 65,535
Minimum:
10
k3
k3,<value2> Returns the current Power Up Permission setting.
k3,<value2> k3,<value2> Sets the Power Up Permission configuration.
<value2>
0 = Operator
1 = Last at Power Off (default)
l => LOAD
l or l0
l0
Begin a load operation.
m => MODE
m or m0 m0,<value2> Returns the current Production Mode.
m0,<value2> m0,<value2> Sets the Production Mode.
<value2>:
0 = Disabled
2 = Dispense (default)
3 = Meter
7 = Dispense minimum chamber volume
8 = Dispense multi-chamber
9 = Meter multi-chamber
m1,<value2> m1,<value2> Sets the Fluidic
Mode. <value2>:
0 = Disabled
1 = Prime (default)
4 = Bubble Clear
6 = Agitate
n => RECIPE
n or n0 n0,<value2> Returns the current Recipe Number.
n0,<value2> n0,<value2> Get the Specified Recipe.
<value2>:
Maximum: 32
Minimum: 1