DIGISPENSE 3020 CONTROLLER MODULE
Page 3-112
7 April, 2011
IVEK Corp.
P03L011B
3.6
PROBLEM GUIDE
Table 3.4 contains a list of possible problems, causes and solutions for the Controller Module.
WARNING
Hazardous voltages exist inside the Controller Module. Under no circumstances should the Controller
Module be opened. There are no user serviceable parts inside the Controller Module. Any unauthorized
access to the inside will void the warranty.
3.7
SPECIFICATIONS
CC TRIG Signal Requirements (Standard):
Accepts mechanical contact closure or solid state switch capable of +5 VDC @ 15mA
(max) Power source in Digispense 3020
DO NOT APPLY VOLTAGE
Logic I/O Trigger In Requirements:
+24 VDC @ 20mA (max), Customer power source
AUX OUT Signal Requirements:
Maximum external voltage: 48 VDC
Maximum current: 250 mA
3.8
MODEL NUMBER
The model number provides important information about the specifics of your Controller Module at time of order. Refer
to this number when calling IVEK Technical support. The model number for your Controller Module is located in the
Title Page section of this manual and on the rear of the Controller Module.
520219 -
# # # # #
Motor
A
- 1200 RPM, 44VDC@3A
Logic Interface
A
- Trig In (CC), Logic (OI)
Parameter Interface
A
- RS232 Serial
Front Panel
A
- Membrane Sw, LCD w/LED Bk Lt
Line Cord & Agency Approval
A
- US Cord & CE
B
- International Cord (no plug) & CE
3.9
ILLUSTRATED PARTS BREAKDOWN
Contact IVEK Corporation Technical Support for information pertaining to replacement parts for this Controller Module.