DIGISPENSE 3020 CONTROLLER MODULE
Page 3-8
7 April, 2011
IVEK Corp.
P03L011B
3.1.7 Pump 90 Degree Offset
Enabling the pump 90 degree offset allows the pump to be offset from the actuator by 90 degrees. This eliminates the
need to mechanically modify the actuator when changing port orientation.
The following table lists the Pump 90 Degree Offset and Active Port settings to change your output from what you
have to what you want. The first table is for the IVEK standard Actuator Module and the second table is for the
optional Actuator Modules. The diagrams at the bottom of the page show the output port locations for the four outlet
positions.
In Forward, Port A is always the input and Port B is the output. In Reverse, Port B is always the input and Port A is
the output.
You
You
Controller Settings:
Have
Want
Pump 90
Active Port
Degree Offset
-03X (std)
-03X
Disabled
Port B
-06X*
Enabled
Port B
-09X
Disabled
Port A
-12X*
Enabled
Port A
You
You
Controller Settings:
Have
Want
Pump 90
Active Port
Degree Offset
-06X
-06X
Disabled
Port B
-09X*
Enabled
Port A
-12X
Disabled
Port A
-06X*
-09X
Enabled
Port B
-09X
Disabled
Port B
-12X*
Enabled
Port A
-06X
-12X
Disabled
Port A
-09X*
Enabled
Port B
-12X
Disabled
Port B
* Rotate the Pump Module 90 Degrees clockwise.
Actuator
Model #
Outlet
Tab
Port
-03X
3 O'clock
-06X
6 O'clock
-09X
9 O'clock
-12X
12 O'clock