DIGISPENSE 3020 CONTROLLER MODULE
Page 3-104
7 April, 2011
IVEK Corp.
P03L011B
Command Response Description
20 = Port Parked
21 = Unparking Port
26 = A fault is being cleared
r => RATE
r or r0
r0,<value2> Returns the current Production
Mode dispense rate
r0,<value2> r0,<value2> Sets the current Production Mode dispense rate in microliters per second.
<value2>:
Maximum:(Pump Dependent)
Minimum: (Pump Dependent)
r1
r1,<value2> Returns the current Fluidic Mode dispense rate.
r1,<value2> r1,<value2> Sets the current Fluidic Mode dispense rate in microliters per second.
<value2>:
Maximum:(Pump Dependent)
Minimum: (Pump Dependent)
r2
r2,<value2> Returns the current Production Mode load rate.
r2,<value2> r2,<value2> Sets the current Production Mode load rate in microliters per second.
<value2>:
Maximum:(Pump Dependent)
Minimum: (Pump Dependent)
r3
r3,<value2> Returns the current Fluidic Mode load rate.
r3,<value2> r3,<value2> Sets the current Fluidic Mode load rate in microliters per second.
<value2>:
Maximum:(Pump Dependent)
Minimum: (Pump Dependent)
s => STATUS
s0
s0,<value2> Returns the volume remaining in the chamber.
s3
s3,<value2> Dump parameters, multiple line response.
s4
s4,<value2> Dump status, multiple line response.
s5
s5,<value2> Returns the Chamber Location - piston location relative to home.
s6
s6,<value2> Returns the Chamber Volume in microliters.
s8
s8,<value2> Returns the current Fault Code.
s9
s9,<value2> Returns the current Error Code.
s10
s10,<value2>Returns the Assert SW Number.
s11
s11,<value2>Returns the Assert SW Code.
s12
s12,<value2>Returns the Assert HW Number.
s13
s13,<value2>Returns the Assert HW Code.
s14
s14,<value2>Returns the Serial Number.
t => DWELL
t2
t2,<value2> Returns the current Pre-Op Dwell in seconds.
t2,<value2> t2,<value2> Sets the Pre-Op Dwell.
<value2>:
Maximum:
2.55
Minimum:
0.00
(default)
t3
t3,<value2> Returns the current Post-Op Dwell in seconds.
t3,<value2> t3,<value2> Sets the Post-Op dwell.
<value2>:
Maximum:
2.55
Minimum:
0.00
(default)