3.2.7.11 Fault Screen
Supervisor
D = Display
P = Push-button Accessible
I/O Test
C = Display/Change N = Not Accessible/Viewable
Operator
0
FAULT
<— status —>
2
3
Piston Home Fault
4
Error Code: ##########
5
6
7
8
9
10
11
“CLEAR FAULT” = CLEAR FAULT & REFERENCE
12
“SETUP A SCREEN” = CHANGE VALUES FIRST
13
If NOT port home fault
14
SETUP A
CLEAR
15
SCREEN
FAULT
If port home fault
14
SETUP A
CLEAR
MOVE
15
SCREEN
FAULT
PISTON
If “MOVE PISTON” selected from above
0
FAULT
<— status —>
2
3
WARNING – trying to move a seized piston
4
may break actuator coupling. Please
5
remove pump before initiating move.
6
7
8
9
10
“MOVE PISTON” = MOVE PISTON & REFERENCE
11
“CLEAR FAULT” = CLEAR FAULT & REFERENCE
12
“SETUP A SCREEN” = CHANGE VALUES FIRST
13
14
SETUP A
CLEAR
MOVE
15
SCREEN
FAULT
PISTON
----- Displays the Controller Module status ........
D D D
----- Displays the fault description ...................... D D D
----- Displays the fault error code ....................... D D D
----- Tests the movement of the piston ............... P P P
----- Clears the displayed fault ........................... P P P
----- Switches to the Setup A screen .................. P P P
----- Displays the Controller Module status ........ D D D
----- Tests the movement of the piston ............... P P P
----- Clears the displayed fault ........................... P P P
----- Switches to the setup A screen .................. P P P