Dispense MCV Minimum/Maximum Chart
Pump Minimum Maximum Minimum Maximum
Size
Volume
Volume
Rate
Rate
µl
µl
µl/Sec
µl/Sec
4A
0.002
200
0.010
40
3A
0.005
500
0.025
100
2A
0.01
1000
0.05
200
1A
0.02
2000
0.10
400
B
0.04
4000
0.20
800
C
0.10
10000
0.50
2000
D
0.20
20000
1.00
4000
Dispense MCV Screen Description -
The Dispense MCV screen displays the current dispense parameters and
allows a user with Supervisor permission to change the parameters. The current permission level determines which
push-buttons are available as shown.
The Start and Stop push-buttons control Dispensing operation while this screen is visible.
Status -
The status indicates the operational state of the system. Refer to section 3.3.9 for an explanation of each
status type.
Total Volume -
Total Volume is the total amount of liquid dispensed in Dispense and Meter Modes. The total volume
counter will roll over after the maximum value is reached. The counter can be reset to 0 using the Clear Totalizer
screen.
Total Cycles -
Total Cycles is the total number of dispenses in Dispense and Meter modes. The total cycles counter
will go up to a maximum of 3,999,999,999 before restarting at 0. The counter can be reset to 0 using the Clear
Totalizer screen.
Recipe -
The Recipe line shows the current recipe selected in the Recipe screen. "0" indicates values changed since
last "Get Recipe".
Pump -
The Pump line shows the current pump size selected in the Setup C screen. The pump size can be either 4A,
3A, 2A, 1A, B, C, D or NONE.
Direction -
The Direction is the direction of liquid flow through the Pump Module for Dispense and Meter modes. The
direction is either Forward or Reverse.
Dispense Volume -
The Dispense Volume is the amount of liquid dispensed in micro liters during a dispense
operation. The range is based on the pump size as shown in the Pump Minimum/Maximum Chart. The increment is
equal to 20% of the minimum volume for the selected pump size.
Dispense/Meter Rate -
The Dispense Rate is the rate of the pump in micro liters per second during the dispense
operation. The range is based on the pump size as shown in the Pump Minimum/Maximum Chart. The increment is
equal to 20% of the minimum rate for the selected pump size.
Load Rate -
The Load Rate is the rate the pump will operate during a load operation in micro liters per second. The
range is based on the pump size as shown in the Pump Minimum/Maximum Chart. The increment is equal to the
minimum rate for the selected pump size.
MCV Conserve Fluid -
MCV Conserve Fluid determines the fluid direction after a reference while moving to the MCV
starting position. The selections are either Disabled (towards the tip) or Enabled (towards the reservoir).
NOTE
Setting MCV Conserve Fluid to enable conserves fluid
by pushing extra fluid back to the reservoir. This may
contaminate the reservoir or cause problems with a
pressurized reservoir.
Drawback -
Refer to section 3.3.8.13.
To change a value;
1. Press the Arrow push-buttons to highlight the
parameter.
2. Press the Enter push-button to bring up either the
Select New Value or Enter New Value screen.
3. In the Select New Value screen use the up and down arrow push-buttons to select the value. In the Enter New
Value screen press the Minimum or Maximum push-buttons to select the desired value or use the left and right arrow
push-buttons to select the digit and the up and down arrow push-buttons to change the digit's value.
4. Press the Enter push-button.