7 | Warning and error messages
INFICON
98 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
Type
Notification
Possible sources of
error
Remedy
W580
Maximum evacuation time
until 100 mbar exceeded
Gross leak at test
object or the
connection to the test
object
• Check the tightness of the connection
between the leak detector and the test
object
• Use another test object, if possible
The settings value for
the max. evacuation
time gross leak is too
low
• Check and change the max. evacuation
time gross leak if necessary
W581
Maximum evacuation time
until measurement
exceeded
Gross leak at test
object or the
connection to the test
object
• Check the tightness of the connection
between the leak detector and the test
object
• Use another test object, if possible
The settings value for
the max. evacuation
time until measurement
is too low
• Check and change the max. evacuation
time until measurement if necessary
W630
Calibration request
Operation mode or
mass has changed
• Perform a calibration
HYDRO•S has been
switched on (UL3000
Fab PLUS only)
• Perform a calibration
Auto Leak Test is
active and testing time
has changed
• Perform a calibration
A new SL3000 sniffer
line has been
connected (UL3000
Fab PLUS only)
• Perform a calibration
Runtime over 30
minutes and
temperature change by
5 Kelvin since last
calibration
• Perform a calibration
Runtime over 30
minutes and no
calibration since 24
hours
• Perform a calibration
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......