6.3.10 Set and use the function ZERO
Why should I use
ZERO?
To be able to clearly measure small leakages the function ZERO should be used.
With each leak detection there is a "Background signal" (see "Definition of terms
8]") that disrupts the search or the measurement of leakages.
• To hide the background signal, activate the function ZERO.
• You can also use ZERO to hide a currently displayed leakage, which interferes
with searches for other or even smaller leaks.
A currently displayed leak is hidden through ZERO.
By using the function ZERO not only is the background signal hidden, but also the
representation of a current leak.
► If you want to prevent this, select the function ZERO only if not measuring a leak at
the same time.
Activate or deactivate
ZERO
ZERO is activated by factory settings.
ü
Supervisor - rights
1
> Setup > Measurement > ZERO and filter
2
If necessary change "ZERO mode". The factory setting is “I•ZERO 2.0”.
ð
If you select "OFF”, the ZERO button does not work.
3
Save
.
ð
If ZERO is activated, start the function by pressing the ZERO button. See also
23]".
How do I switch the
ZERO function ON?
Briefly press the ZERO button on the control panel or the touchscreen
.
How do I switch the
ZERO function OFF?
Press the ZERO button on the control panel for more than 2 seconds or on the
touchscreen
.
Operation mode
“Vacuum”, effect on
ZERO
After pressing the ZERO button, the displayed background signal is dimmed as a
function of the current drift. Typically, the display limit falls in several steps. The size of
the current display limit is a dimension with which leakages can currently be
measured. When measuring, the current display limit should be at least one decade
below the expected measured value.
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......