10 | Accessories and interfaces
INFICON
120 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
10.2.4 Configuring digital outputs
The digital outputs PLC-OUT 1 ... The available functions can be assigned in any way
necessary to the 8 I/O module.
ü
Supervisor - rights
1
> Setup > Accessories > I/O module > Digital outputs configuration
2
Select the desired "Digital output".
ð
The outputs PLC-OUT 1 - 8 are available.
3
Select the desired "Function", see the table below.
4
Select the desired "Mode".
ð
The modes "Standard" or the mode "Inverse" are available for selection.
To mode "Inverse": In the table overview the description of the state "open"
must be replaced with the description of the state "closed".
5
Save
.
Function
State:
Description
Off (open)
open:
always open
Ready for operation
closed:
Emission switched on, calibration process
inactive, no error
open:
Emission switched off or calibration process
active or error
Emission on
closed:
Emission switched on
open:
Emission switched off
Error
closed:
Error
open:
No error
Error or warning
closed:
Error or warning
open:
No error or warning
Run-up
closed:
Run up
open:
no run-up
Calibration request
closed:
No external calibration: Calibration required
When external calibration or "CAL check":
Request "Open or close external calibrated
leak"
open:
No request: Calibration not required
Calibration active
closed:
Device is to be calibrated.
open:
Device is not to be calibrated.
Calibration stable
closed:
Calibration completed with calibration leak
open:
Assignment not stable or calibration is
inactive
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......