8 | Cleaning and maintenance
INFICON
106 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
Table 3: Capillary filter
1 Sniffer line end
4 Plastic capillary filter
2 cone seal
5 Water conservation tip
3 Metal capillary filter
Capillary filter and water conservation tip are screwed onto the sniffer line end. The
filter inserts are inserted and lie in front of a small metal grill that is also inserted.
To replace the filter inserts, proceed as follows:
1
Switch off the leak detector.
2
Unscrew the capillary filter or the water conservation tip. Make sure that the
cone seal does not fall out.
3
Push out the old filter pads and the metal grill from the rear.
Fig. 17: Pressing the filter inserts out of the capillary filter
4
Dispose of the old inserts and clean the metal grill.
5
Push the metal grill and then the two new inserts into the filter from the front.
Make sure that grill and inserts do not cant.
6
Switch on the leak detector.
7
Close the sniffer probe with your finger. With the water protection tip, you have
to also cover the opening on the side. You should then be able to feel a
negative pressure. If not then there is a leak somewhere and you have to check
the screw connection. The cone seal may also have fallen out.
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......